DA-1

Technical Parameters

Basic parameters

Model DA-1
Input Power DC5V 1A
Batteries 3.7V/1400mAh
Tip amplitude minimum:1μm, deviation -50%
maximum:100μm, deviation +50%
Half deflection force minimum:0.1 N, deviation -50%
maximum:5N,deviation +50%
Vibration frequency of tip 18kHz~45kHz
Output Power of tip 3W MAX
Weight of main unit 116g
main unit size 170mm*25mm*50mm
Packing Size 200mm*125mm*60mm
Operation modes discontinuous running
Storage Environment Temperature:-20℃~55℃
Relative humidity:≤90%
Atmospheric pressure:70kPa~106kPa
Working Environment Temperature:5℃~40℃
Relative humidity: ≤80%
Atmospheric pressure: 70kPa~106kPa
Life time 5 year
Tip/Wrench/Silicone sleeve can be sterilized 134℃ and 0.22MPa pressure
 
 

Accessories

Videos

F&Q

  • DA-1 Maintenance Guide

    DA-1 Maintenance Guide

    I. Daily Maintenance

    1.1 Before each use, sterilize the working tip, silicone sleeve, and wrench accessories in a steam sterilizer at 134°C and 0.22 MPa.

    1.2 After treatment, use the wrench to remove the working tip. Wipe the working tip, silicone sleeve, and wrench with an alcohol swab to remove any blood residues.

    1.3 After treatment, turn off the power promptly. Use a clean cloth dampened with water to wipe the surface of the device.

    II. Periodic Maintenance

    2.1 Replace the working tip regularly. A damaged or heavily worn tip may reduce vibration performance. It is recommended to replace the tip after use in no more than 20 root canals.

    2.2 Replace the silicone sleeve regularly. The recommended replacement cycle is every 6 months.

    III. Sterilization Instructions

    3.1 Disinfect the working tip, silicone sleeve, and wrench with alcohol. Soak a clean dry gauze in 70%-80% ethanol, then wipe the surfaces to be disinfected twice. Allow 3 minutes for disinfection time. Let air dry naturally or use a clean, dry cloth to remove any remaining disinfectant.

    3.2 Sterilize the working tip, silicone sleeve, and wrench in a high-temperature, high-pressure steam sterilizer at 134°C, 2.2 bar (0.22 MPa) for 4 minutes.

    3.3 The maximum number of sterilization cycles is 250 times.

    3.4 The following disinfection methods are strictly prohibited:

    Boiling in any liquid solution

    Soaking in iodine, alcohol, glutaraldehyde, or other disinfectant solutions

    Heating in ovens or microwave ovens

    IV. Precautions for Use

    4.1 Do not attach or detach the working tip while it is vibrating.

    4.2 The working tip must be tightly secured before use.

    4.3 Do not bend or grind the working tip.

    4.4 Stop using the device if a harsh whistling sound occurs during use.

    4.5 Do not apply excessive pressure when performing root canal irrigation.

    V. Protection and Storage

    5.1 Handle and store the product with care. Keep it away from sources of vibration and store it in a cool, dry, well-ventilated environment.

    5.2 Do not store the product with toxic, corrosive, flammable, or explosive materials.

    5.3 Store the product in an environment with a temperature between -20°C and 55°C, relative humidity between 0% and 90%, and atmospheric pressure between 70 kPa and 160 kPa.